Article ID Journal Published Year Pages File Type
5498313 Physica Medica 2017 9 Pages PDF
Abstract
PD at the thyroid with Iris™ was about a third lower than with a fixed collimator in case of brain SRT. Despite uncertainties (use of anthropomorphic PD to estimate patient specific PD, surface PD to estimate OAR PD) the model allows PD to be estimated without in vivo measurements. This method could be used to optimise PD with different planning strategies.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , , , , , , , ,