Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5498347 | Physica Medica | 2017 | 7 Pages |
Abstract
We validated the dose reduction capability of additional Cu-filters without degrading the image quality for abdominal radiography. The estimated entrance surface dose reductions of the Cu-filters were approximately 30-40% and 20-30% for the acrylic and bone contrasts, respectively, and effective dose reductions for acrylic were nearly half of those for ESD. At these reduced dose conditions, the current time product values needed to be increased by factors of 1.4 and 1.8 for the 0.1- and 0.2-mm-thick Cu-filters, respectively.
Keywords