Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5498362 | Physica Medica | 2016 | 4 Pages |
Abstract
Results indicate a difference of about â32% in voxel S factors for 111In in 4.42Â mm voxel size and around â35% in 4.80Â mm voxel size, particularly self-dose values; this lead to significant shift in dose histograms and average doses. For 177Lu and 90Y, differences are about 2% and 12% for 4.42Â mm voxels and about â8% and 9% for 4.80Â mm voxels, respectively, attributable to the different calculation methods of the voxel S factors; this does not lead to significant discrepancies between the two dose histograms. Consequently, voxel S factors must account accurately for all radiations emitted by the nuclide.