Article ID Journal Published Year Pages File Type
5498543 Physica Medica 2017 7 Pages PDF
Abstract
Fast, epithermal and thermal photoneutron dosimetry of 6 MV X-rays were made by novel dosimetry methods in a Siemens ONCOR accelerator with sum dose equivalent per Gy of ∼0.0014% μSv with ∼0.21 MeV mean energy at the isocenter; i.e. ∼150 times smaller than that of 18 MV X-rays. This observation assures clinical safety of 6 MV X-rays in particular in single-mode machines like Elekta COMPACT producing no photoneutrons due to no “beryllium exit window” in the head structure.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation