Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5498589 | Physica Medica | 2017 | 9 Pages |
Abstract
Dosimetry in small radiation field is challenging and complicated because of dose volume averaging and beam perturbations in a detector. We evaluated the suitability of the “Edge-on” MOSkin (MOSFET) detector in small radiation field measurement. We also tested the feasibility for dosimetric verification in stereotactic radiosurgery (SRS) and stereotactic radiotherapy (SRT). “Edge-on” MOSkin detector was calibrated and the reproducibility and linearity were determined. Lateral dose profiles and output factors were measured using the “Edge-on” MOSkin detector, ionization chamber, SRS diode and EBT2 film. Dosimetric verification was carried out on two SRS and five SRT plans. In dose profile measurements, the “Edge-on” MOSkin measurements concurred with EBT2 film measurements. It showed full width at half maximum of the dose profile with average difference of 0.11 mm and penumbral width with difference of ±0.2 mm for all SRS cones as compared to EBT2 film measurement. For output factor measurements, a 1.1% difference was observed between the “Edge-on” MOSkin detector and EBT2 film for 4 mm SRS cone. The “Edge-on” MOSkin detector provided reproducible measurements for dose verification in real-time. The measured doses concurred with the calculated dose for SRS (within 1%) and SRT (within 3%). A set of output correction factors for the “Edge-on” MOSkin detector for small radiation fields were derived from EBT2 film measurement and presented. This study showed that the “Edge-on” MOSkin detector is a suitable tool for dose verification in small radiation field.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
Wei Loong Jong, Ngie Min Ung, Ath Vannyat, Zulaikha Jamalludin, Anatoly Rosenfeld, Jeannie Hsiu Ding Wong,