Article ID Journal Published Year Pages File Type
5498592 Physica Medica 2017 7 Pages PDF
Abstract
The AC with Multiplug allowed for comprehensive commissioning of CyberKnife MC algorithm and is useful for patient specific QA for stereotactic body radiation therapy. MC calculation accuracy might be limited due to Multiplan's insufficient material library; still results are comparable with other reported commissioning measurements using FFF beams.
Keywords
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , , , , , , , ,