Article ID Journal Published Year Pages File Type
5517056 Biologicals 2017 6 Pages PDF
Abstract

A size-exclusion high-performance liquid chromatographic method using a methacrylate-based column was developed, validated and implemented for the determination of pegfilgrastim aggregates. The samples were directly injected into a TSKgel G4000PWXL column (7.5 mm × 300 mm, 10 μm, <500 A°) with a mobile phase of 100 mM phosphate, pH 2.5. Detection was made at 215 nm and analyses were run at a flow-rate of 0.6 ml/min at 10 °C. Vortex-mixing of samples produced oligomers, however, very high molecular weight aggregates were formed at high temperatures. The method exhibited linearity over the concentration range of 0.1-14 mg/ml for pegfilgrastim monomer and high molecular weight aggregates with a correlation coefficient of greater than 0.99. The method was specific and sensitive, with a lower quantification limit of 0.1 mg/ml and a detection limit of 0.02 mg/ml. Over 1200 samples were analyzed by the present method without significant change in the column performance.

Related Topics
Life Sciences Biochemistry, Genetics and Molecular Biology Biochemistry, Genetics and Molecular Biology (General)
Authors
, , , , , ,