Article ID Journal Published Year Pages File Type
5575757 The Foot 2017 5 Pages PDF
Abstract
There is a high rate of complications with the Toefit-Plus™ implant resulting in revision surgery. Patients should have the risks associated with arthroplasty clearly explained, including the risk of revision, and the option of arthrodesis should be discussed when planning surgery. Further trials and redesign of implants may help to improve results. The authors would not recommend the Toefit-Plus™ implant due to poor results seen in one third of patients.
Keywords
Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation