Article ID Journal Published Year Pages File Type
5597461 The Annals of Thoracic Surgery 2017 6 Pages PDF
Abstract
The EVAHEART LVAS caused significantly less vWF degradation than the HeartMate II in a mock circulatory loop with whole human blood. LVAD design features may minimize vWF degradation. These data may inform the design and operation of next-generation LVADs to minimize blood trauma.
Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine