Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5610238 | Journal of the American Society of Echocardiography | 2013 | 9 Pages |
Abstract
This study establishes parameters (mean ± 2 SDs) defining the distribution of values for Doppler-derived hemodynamic data with normal St. Jude Medical mechanical mitral valve prostheses. Prostheses with hemodynamic values outside these parameters are likely dysfunctional; however, prosthesis dysfunction may be present even when hemodynamic values are within these ranges.
Keywords
PHTTTETEELVOTEOAGOALVEFiEOAEAESJMTVIppmPPIASESVIBSATime-velocity integralAmerican Society of EchocardiographyEuropean Association of EchocardiographyechocardiographyTransthoracic echocardiographyTransesophageal echocardiographystroke volumeDopplermitral valveheart valvestwo-dimensionalbody surface areaSt. Jude Medicalstroke volume indexpressure half-timeleft ventricular outflow tractgeometric orifice areaindexed effective orifice areaeffective orifice areaProsthesisprosthesis-patient mismatchleft ventricular ejection fractionmean gradient
Related Topics
Health Sciences
Medicine and Dentistry
Cardiology and Cardiovascular Medicine
Authors
Lori A. MD, Joseph F. MD, Heidi M. MD, David O. MS, Regina M. BS, Rakesh M. MD, DPhil, Fletcher A. MD,