Article ID Journal Published Year Pages File Type
5610939 Journal of the American Society of Echocardiography 2009 4 Pages PDF
Abstract
Cardiac perforation and device erosion are infrequent but potentially lethal complications of percutaneous occluder device insertion. Transthoracic and transesophageal echocardiography (TEE) and gated cardiac computed tomography are useful in detecting this complication. In particular, TEE can detect characteristic features to confirm the diagnosis before rapid surgical correction. Deficient superior-anterior rim and encroachment of the occluder device on the posterior atrial wall are risk factors for device erosion. TEE performed before device insertion with awareness of manufacturer's guidelines for atrial geometry and device sizing may reduce the occurrence of this serious complication.
Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine
Authors
, , , , , ,