Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5611057 | Journal of the American Society of Echocardiography | 2009 | 6 Pages |
Abstract
Normal Doppler-derived echocardiographic data for Carpentier-Edwards Duraflex (Edwards Lifesciences, Irvine, CA) porcine bioprosthesis function in the mitral position are limited to 2 small series that did not include all Doppler-derived variables. The purpose of this study was to provide a comprehensive Doppler echocardiographic assessment of normal Carpentier-Edwards Duraflex mitral bioprosthesis function in a large number of patients assessed in the early postoperative phase. All of the important Doppler-derived hemodynamic variables reported to date were used. All patients had either a mitral valve prosthesis time velocity integral to left ventricular outflow tract time velocity integral ratio < 3.9 or an E velocity < 2.8. The pressure half-time was < 130 msec in all patients. Nearly all patients (97%) had an E velocity < 2.8 msec and a mitral valve prosthesis time velocity integral to left ventricular outflow tract time velocity integral ratio < 3.9 regardless of bioprosthesis size, left ventricular function, heart rate, hemoglobin, or hematocrit. With increasing bioprosthesis size, effective orifice area significantly increased, whereas the prosthesis performance index significantly decreased.
Keywords
Related Topics
Health Sciences
Medicine and Dentistry
Cardiology and Cardiovascular Medicine