Article ID Journal Published Year Pages File Type
5664572 Transfusion and Apheresis Science 2017 4 Pages PDF
Abstract

Peripheral blood progenitor cells (PBPC) infusion allows rapid haematological recovery after high dose chemotherapy. Efficient PBPC collection is therefore essential as rescue therapy for transplantation. In order to validate a new equipment (ComTec®, Fresenius Kabi), we compared the efficiency of three cell separators for PBPC collection in patients with haematological malignant diseases. From June 2014 to December 2015, 83 PBPC were collected in 48 patients. Three aphaeresis machines were used: Cobe Spectra® (Terumo BCT, 11), Amicus® (Fenwall, 30), and ComTec® (Fresenius Kabi, 42). The median collection efficiency was similar between the three separators. The evaluation of cell contamination in the final product revealed a lower red cell contamination with Spectra® and ComTec®, whereas the platelet contamination was lower with Amicus®. The new equipment has been validated and can be further used in routine, with a total running cost that turned out to be quite lower. Each separator has its own characteristics and advantages. Further study is needed to suggest that the choice of separator could be guided following the patient's blood characteristics.

Related Topics
Health Sciences Medicine and Dentistry Hematology
Authors
, , , , , , ,