This is the first report of a rare occurrence of intraocular and extraocular hemorrhage associated following spontaneous release of ligature of a non-valved glaucoma drainage implant. The presumed mechanism was sudden shallowing of the anterior chamber resulting in the tube irritating uveal vasculature. We do not have an explanation for the extraocular blood.