Article ID Journal Published Year Pages File Type
5707666 Gait & Posture 2017 5 Pages PDF
Abstract
The in-situ PFA calibration was performed on six plates by assessing the locations of up to 98 points of force application per plate using an instrumented pole. Application of the in-situ PFA calibration approach reduced the root mean square errors by up to approximately 60% compared to the manufacturers calculation. Correction coefficients were strongly dependent on the individual platform as well as the location of the applied force on the plate. Remounting of the plates altered the plate coefficient corrections considerably, while changes over time were notable but not as extensive. We therefore recommend that plates should be recalibrated after remounting, but also at least every 5 years, in order to ensure the preservation of an accurate PFA.
Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, , , , ,