Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5708531 | The Journal of Arthroplasty | 2017 | 29 Pages |
Abstract
In our series of the recalled LFIT CoCr femoral head, the risk of impending GTF or head-neck taper corrosion should be considered as a potential diagnosis in a painful LFIT femoral head and Accolade titanium-molybdenum-zirconium-iron alloy THA with unknown etiology. Almost half of the failures we observed included sizes outside of the voluntary recall.