Prophylactic cerclage wire placement increases the rotation and energy to failure in well-fixed press-fit femoral implants. The increase in torsional energy needed for failure may reduce the risk of early periprosthetic fracture. Further studies are needed to evaluate cost vs benefit and long-term outcomes of prophylactic wiring. Based on the results of our study, consideration of prophylactic wiring should be addressed on a case-to-case basis.