5726381 European Journal of Radiology 2017 4 Pages PDF
Abstract

•This study aimed to determine optimal scanning parameters for commonly-used tumor ablation applicators.•The findings illustrate the overall interaction of the effects of kVp, ASiR, and reconstruction algorithm within and between probes, so that radiologists may easily reference optimal imaging performance.•Optimum combinations for each probe are provided.

PurposeCT-beam hardening artifact can make tumor margin visualization and its relationship to the ablation applicator tip challenging. To determine optimal scanning parameters for commonly-used applicators.Materials and methodsApplicators were placed in ex-vivo cow livers with implanted mock tumors, surrounded by bolus gel. Various CT scans were performed at 440 mA with 5 mm thickness changing kVp, scan time, ASiR, scan type, pitch, and reconstruction algorithm. Four radiologists blindly scored the images for image quality and artifact quantitatively.ResultsA significant relationship between probe, kVp level, ASiR level, and reconstruction algorithm was observed concerning both image artifact and image quality (both p = <0.0001). Specifically, there are certain combinations of kVp, ASiR, and reconstruction algorithm that yield better images than other combinations. In particular, one probe performed equivalently or better than any competing probe considered here, regardless of kVp, ASiR, and reconstruction algorithm combination.ConclusionThe findings illustrate the overall interaction of the effects of kVp, ASiR, and reconstruction algorithm within and between probes, so that radiologists may easily reference optimal imaging performance for a certain combinations of kVp, ASiR, reconstruction algorithm and probes at their disposal. Optimum combinations for each probe are provided.

Related Topics
Health Sciences Medicine and Dentistry Radiology and Imaging
Authors
, , , , , , , ,