5726435 European Journal of Radiology Open 2017 6 Pages PDF
Abstract

•Quantitative Perfusion CT parameters are not affected substantially by increasing ASIR levels.•ASIR improves image noise and CNR at a reduced peak kilovoltage of 100 kV.•Peak tumour CNR is greater than 1.5 with 60% ASIR and above.

ObjectivesTo determine the effect of Adaptive Statistical Iterative Reconstruction (ASIR) on perfusion CT (pCT) parameter quantitation and image quality in primary colorectal cancer.MethodsProspective observational study. Following institutional review board approval and informed consent, 32 patients with colorectal adenocarcinoma underwent pCT (100 kV, 150 mA, 120 s acquisition, axial mode). Tumour regional blood flow (BF), blood volume (BV), mean transit time (MTT) and permeability surface area product (PS) were determined using identical regions-of-interests for ASIR percentages of 0%, 20%, 40%, 60%, 80% and 100%. Image noise, contrast-to-noise ratio (CNR) and pCT parameters were assessed across ASIR percentages. Coefficients of variation (CV), repeated measures analysis of variance (rANOVA) and Spearman' rank order correlation were performed with statistical significance at 5%.ResultsWith increasing ASIR percentages, image noise decreased by 33% while CNR increased by 61%; peak tumour CNR was greater than 1.5 with 60% ASIR and above. Mean BF, BV, MTT and PS differed by less than 1.8%, 2.9%, 2.5% and 2.6% across ASIR percentages. CV were 4.9%, 4.2%, 3.3% and 7.9%; rANOVA P values: 0.85, 0.62, 0.02 and 0.81 respectively.ConclusionsASIR improves image noise and CNR without altering pCT parameters substantially.

Related Topics
Health Sciences Medicine and Dentistry Radiology and Imaging
Authors
, , , , , , ,