Abstract
An original flow injection (FI) system was developed for on-line microextraction of Vanadium (V) based on room temperature ionic liquid (RTIL). Vanadium was complexed with 2-(5-bromo-2-pyridylazo)-5-diethylaminophenol (5-Br-PADAP) at pH 4.0. A 40 μL-volume of 1-butyl-3-methylimidazolium hexafluorophosphate ([C4mim][PF6]) RTIL was mixed with 5 mL of sample solution containing the V-5-Br-PADAP complex. Then, a fully on-line temperature-assisted dispersion procedure was developed, followed by, analyte microextraction; and final on-line separation of the RTIL phase with a florisil-containing microcolumn. Vanadium was removed from the microcolumn with a 10% (v/v) nitric acid (in acetone) solution, and finally measured by electrothermal atomic absorption spectrometry (ETAAS). The detection limit achieved after preconcentration of 5 mL of sample solution, was 4.8 ng Lâ1. The relative standard deviation (RSD) for 10 replicate determinations at 5 μg Lâ1 of vanadium level was 4.1%, calculated from the obtained peak heights. The calibration graph was linear, with a correlation coefficient of 0.9982 at levels from the detection limits up to 15 μg Lâ1. The method was successfully applied for the determination of vanadium in environmental and biological samples.
Keywords
Related Topics
Physical Sciences and Engineering
Chemical Engineering
Chemical Health and Safety
Authors
Paula Berton, EstefanÃa M. Martinis, Rodolfo G. Wuilloud,