Article ID Journal Published Year Pages File Type
580701 Journal of Hazardous Materials 2010 7 Pages PDF
Abstract
The on-line incorporation of cloud point extraction (CPE) with/without 8-hydroxyquinoline (8-Ox) as chelating agent into flow injection analysis associated with inductively coupled plasma optical emission spectrometry (ICP-OES) for determining trace rare earth elements (REEs) is presented and evaluated. The significant parameters affecting on-line cloud point extraction of REEs such as sample pH, flow rate, 8-Ox concentration, Triton X-114 concentration were systematically studied. Under the optimized conditions, with the consumption of 3.0 mL sample solution, the limits of detection (3σ) were ranged from 41.4 pg mL−1 (Yb) to 448 pg mL−1 (Gd) with relative standard deviations (RSDs) of 1.0% (Eu)-5.9% (Sm) for on-line CPE-ICP-OES with 8-Ox as chelating agent, and 69.0 pg mL−1 (Sc) to 509.5 pg mL−1 (Sm) with RSDs of 2.9% (Yb)-7.5% (Ho) for on-line CPE-ICP-OES without 8-Ox as chelating agent, respectively. The sample throughput of 17 samples h−1 was obtained for both systems. The developed methods of on-line CPE-ICP-OES were validated by the analysis of certified reference material (GBW07605, tea leaves) and real biological samples of pig liver, Auricularia auricula and mushroom.
Related Topics
Physical Sciences and Engineering Chemical Engineering Chemical Health and Safety
Authors
, ,