Abstract
A new method based on the cloud point extraction (CPE) preconcentration and graphite furnace atomic absorption spectrometry (GFAAS) detection was proposed for the determination of trace nickel in water samples. When the micelle solution temperature is higher than the cloud point of surfactant p-octylpolyethyleneglycolphenyether (Triton X-100), the complex of Ni2+ with 1-phenyl-3-methyl-4-benzoyl-5-pyrazolone (PMBP) could enter surfactant-rich phase and be concentrated, then determined by GFAAS. The main factors affecting the cloud point extraction were investigated in detail. An enrichment factor of 27 was obtained for the preconcentration of Ni2+ with 10 mL solution. Under the optimal conditions, the detection limit of Ni2+ is 0.12 ng mLâ1 with R.S.D. of 4.3% (n = 10, c = 100 ng mLâ1). The proposed method was applied to determination of trace nickel in water samples with satisfactory results.
Related Topics
Physical Sciences and Engineering
Chemical Engineering
Chemical Health and Safety
Authors
Zhimei Sun, Pei Liang, Qiong Ding, Jing Cao,