Article ID Journal Published Year Pages File Type
5863315 International Emergency Nursing 2016 6 Pages PDF
Abstract

•New blood control IV catheter was compared to a standard without blood control.•Similar clinical acceptability and insertion success rates of the 2 IV catheters.•Blood control catheter significantly decreased clinician and patient blood exposure.

IntroductionThe performance of a new safety peripheral intravenous catheter (PIVC) that contains a blood control feature in the hub (blood control) was compared against the current hospital standard without blood control (standard).MethodsIn this prospective, non-blinded trial, patients were randomized 1:1 to receive either device. Insertions were performed and rated by emergency room nurses. Primary endpoints included clinical acceptability, incidence of blood leakage, and risk of blood exposure. Secondary endpoints were digital compression, insertion success, and usability.Results15 clinicians performed 152 PIVC insertions (73 blood control, 79 standard). Clinical acceptability of the blood control device (100%) was non-inferior to the standard (98.7%) (p < 0.0001). The blood control device had a lower incidence of blood leakage (14.1% vs 68.4%), was superior in eliminating the risk of blood exposure (93.9% vs 19.1%) and the need for digital compression (95.3% vs 19.1%), while maintaining non-inferior insertion success rates (95.9% vs 93.7%) and usability ratings (p < 0.0001).DiscussionIn comparison with the hospital-standard, the new safety PIVC with integrated blood control valve had similar clinical acceptability ratings yet demonstrated superior advantages to both clinicians and patients to decrease blood leakage and the clinician's risk of blood exposure, during the insertion process.

Related Topics
Health Sciences Medicine and Dentistry Critical Care and Intensive Care Medicine
Authors
, , , , , ,