Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5872343 | Clinical Nutrition | 2013 | 5 Pages |
Abstract
Despite strong correlations and small mean biases for VO2, VCO2 and REE obtained by the Vmax® and E-COVX®, the limits of agreement were beyond the clinically acceptable range. These devices should not be used interchangeably for gas exchange measurements in mechanically ventilated children.
Related Topics
Health Sciences
Medicine and Dentistry
Critical Care and Intensive Care Medicine
Authors
Craig D. Smallwood, Nilesh M. Mehta,