Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5984472 | Journal of Cardiology Cases | 2015 | 4 Pages |
In Japan, transcatheter aortic valve implantation (TAVI) with Edwards-SAPIEN XT valve (Edwards Lifesciences Inc., Irvine, CA, USA) started in October 2013. All institutions should undergo a training period to perform TAVI independently. Balloon aortic valvuloplasty (BAV) as a bridge to TAVI during the training period should be performed with caution to avoid severe aortic regurgitation (AR) because bailout TAVI is not possible. We present a case in which BAV was successfully performed as a bridge to TAVI during the training period. The patient was an 85-year-old man with medically uncontrollable congestive heart failure due to severe aortic valve stenosis. The aortic valve area was 0.60Â cm2 with a left ventricular ejection fraction of 20%. TAVI was considered a safe but high-risk strategy owing to the unstable hemodynamic condition. We chose BAV as a bridge therapy to TAVI. The aortic annulus diameter was 25.3Â mm on computed tomography scans. We chose a 20-mm balloon catheter to avoid BAV-induced AR. Transfemoral TAVI was performed successfully 16 days after BAV using a 26-mm SAPIEN XT valve. The postoperative course was uneventful. The case demonstrated BAV as a bridge therapy to TAVI can be safely and effectively performed during the training period.