Abstract

AimAccurate chest compression detection is key to evaluate cardiopulmonary resuscitation (CPR) quality. Two automatic compression detectors were developed, for the compression depth (CD), and for the thoracic impedance (TI). The objective was to evaluate their accuracy for compression detection and for CPR quality assessment.MethodsCompressions were manually annotated using the force and ECG in 38 out-of-hospital resuscitation episodes, comprising 869 min and 67,402 compressions. Compressions were detected using a negative peak detector for the CD. For the TI, an adaptive peak detector based on the amplitude and duration of TI fluctuations was used. Chest compression rate (CC-rate) and chest compression fraction (CCF) were calculated for the episodes and for every minute within each episode. CC-rate for rescuer feedback was calculated every 8 consecutive compressions.ResultsThe sensitivity and positive predictive value were 98.4% and 99.8% using CD, and 94.2% and 97.4% using TI. The mean CCF and CC-rate obtained from both detectors showed no significant differences with those obtained from the annotations (P > 0.6). The Bland-Altman analysis showed acceptable 95% limits of agreement between the annotations and the detectors for the per-minute CCF, per-minute CC-rate, and CC-rate for feedback. For the detector based on TI, only 3.7% of CC-rate feedbacks had an error larger than 5%.ConclusionAutomatic compression detectors based on the CD and TI signals are very accurate. In most cases, episode review could safely rely on these detectors without resorting to manual review. Automatic feedback on rate can be accurately done using the impedance channel.

Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine
Authors
, , , , , , ,