Article ID Journal Published Year Pages File Type
6003407 Thrombosis Research 2007 12 Pages PDF
Abstract

IntroductionThe aim of the study was to evaluate the test characteristics of the Platelet Function Analyzer-100 (PFA-100) in patients treated with aspirin.Methods and resultsThe study consisted of two sub-studies. In study 1, 10 patients with ischemic heart disease (IHD) and 10 controls had platelet function assessed by optical platelet aggregation and the PFA-100 method in two 5-week periods. Patients with IHD were treated with aspirin 150 mg/day (first 5-week period), and 300 mg/day (second 5-week period), whereas the controls only received aspirin (150 mg/day) during the second 5-week period. From the results of study 1, we found that a cut-off value for the PFA-100 collagen/epinephrine cartridge < 165 s identified patients not taking aspirin (sensitivity 0.91, specificity 1.00). A good agreement between the PFA-100 method and optical platelet aggregation was found. Within-subject variation for the PFA-100 collagen/epinephrine cartridge was ± 28%, as compared to ± 17% for the optical platelet aggregation. Study 2 included 298 aspirin treated patients who were admitted with symptoms suggestive of an acute myocardial infarction. Platelet function was assessed in duplicate by the PFA-100 collagen/epinephrine cartridge, and a 95% Limit of Agreement interval at [− 65%, 65%] indicated a limited precision.ConclusionWe defined a cut-off value below which patients not taking aspirin can be identified. However, due to imprecision of the PFA-100 method repeated duplicate assessment of the collagen/epinephrine Closure Time is recommended.

Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine
Authors
, , , , , ,