Article ID Journal Published Year Pages File Type
6037385 NeuroImage 2010 12 Pages PDF
Abstract
In this work, a B1 mapping method based on magnetization-prepared FLASH with slice selective preparation and excitation pulses and correction for relaxation effects is presented, yielding B1 maps with whole brain coverage, an in-plane resolution of 4 mm, a slice thickness of 3 mm, and a clinically acceptable duration of 46 s. The method is tested both in vitro and in vivo and applied in a subsequent in vivo study to show that MTR values in human brain tissue depend approximately linearly on the preparation angle, with a slope similar to values reported for 1.5 T. Calibration data and B1 maps are applied to B1 inhomogeneity corrections of MTR maps. Subsequently, it is shown that B1-corrected MTR maps acquired at reduced preparation angles due to individual SAR restrictions can be normalized, allowing for a direct comparison with maps acquired at the full angle.
Related Topics
Life Sciences Neuroscience Cognitive Neuroscience
Authors
, , , ,