Article ID Journal Published Year Pages File Type
6045077 World Neurosurgery 2015 6 Pages PDF
Abstract
These data suggest that the incorporation of the Ellman Surgitron Generator can result in the reduction of thermal transfer with conventional bipolar forceps compared with other generators. The combination with Codman ISOCOOL forceps can maximize the potential safety associated with bipolar coagulation. With regard to the use of comarketed pairs of forceps and generators, the combination of Ellman Surgitron Generator and Ellman bipolar forceps provided the best thermal profile.
Related Topics
Life Sciences Neuroscience Neurology
Authors
, , , , , ,