Article ID Journal Published Year Pages File Type
6079146 The American Journal of Emergency Medicine 2015 5 Pages PDF
Abstract

PurposeFeedback devices are used to improve chest compression (CC) quality related to survival rates in cardiac arrest. However, several studies have shown that feedback devices are not sufficiently reliable to ensure adequate CC depth on soft surfaces. Here, we determined the proper target depth of feedback (TDF) using an accelerometer during cardiopulmonary resuscitation in hospital beds.MethodsIn prospective randomized crossover study, 19 emergency physicians performed CCs for 2 minutes continuously on a manikin in 2 different beds with 3 TDFs (5, 6, and 7 cm). We measured CC depth, the proportion of accurate compression depths, CC rate, the proportion of incomplete chest decompressions, the velocity of CC (CC velocity), the proportion of time spent in CC relative to compression plus decompression (duty cycle), and the time spent in CC (CC time).ResultsMean (SD) CC depths at TDF 5, 6, and 7 were 45.42 (5.79), 52.68 (4.18), and 58.47 (2.48) on one bed and 46.26 (4.49), 53.58 (3.15), and 58.74 (2.10) mm on the other bed (all P < .001), respectively. The proportions of accurate compression depths and CC velocity at TDF 5, 6, and 7 differed significantly according to TDF on both beds (all P < .001).The CC rate, CC time, and proportion of incomplete chest decompression did not differ on both beds (all P > .05). The duty cycle differed significantly on only B2.ConclusionsThe target depth of the real-time feedback device should be at least 6 cm but should not exceed 7 cm for optimal CC on patients on hospital beds.

Related Topics
Health Sciences Medicine and Dentistry Emergency Medicine
Authors
, , , , , , , , ,