The aim of this study was to compare the diagnostic performance of 2 interferon-γ release assays, an enzyme-linked immunospot assay (T-SPOT.TB; Oxford Immunotec Ltd., Oxford, UK) and the QuantiFERON-TB Gold in-Tube assay (QFT-GIT; Cellestis Ltd., Carnegie, Australia), in patients with suspected active tuberculosis (TB). From October 2009 to October 2011, a total of 200 patients with suspected TB were enrolled. Clinical and microbiological characteristics of the patients were collected and blood samples were obtained for T-SPOT.TB and QFT-GIT assays. Among the 200 subjects, 98 (49%) had culture-confirmed TB, 18 (9%) had probable TB, and the remaining 84 (42%) subjects did not have TB. The sensitivity, specificity, positive predictive value, and negative predictive value for active TB diagnosis by the T SPOT. TB were 83%, 71%, 81%, and 75%, respectively. For QFT-GIT, the sensitivity, specificity, positive predictive value, and negative predictive value for active TB diagnosis were 66%, 76%, 80%, and 62%, respectively. The QFT-GIT assay resulted in more indeterminate and false-negative results than the T-SPOT.TB assay, especially in immunocompromised patients. In conclusion, T-SPOT.TB had a higher sensitivity and resulted in fewer indeterminate results than the QFT-GIT assay for diagnosing active TB.