Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
6132882 | Journal of Virological Methods | 2016 | 8 Pages |
Abstract
The linear dynamic range of each of the four assays spanned eight orders of magnitude (104-1011 copies/mL). The limit of detection for assays 1-4 was estimated to be 1147 copies/mL, 765 copies/mL, 1138 copies/mL and 1470 copies/mL respectively. Each assay demonstrated a strong linear relationship (r2Â =Â >0.995) and amplification efficiency greater than 95%. Repeatability and reproducibility of the method were shown to be high, with coefficients of variations lower than 8% and 15% respectively.
Related Topics
Life Sciences
Immunology and Microbiology
Virology
Authors
Chisha T. Sikazwe, Glenys R. Chidlow, Allison Imrie, David W. Smith,