Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
6133386 | Journal of Virological Methods | 2015 | 22 Pages |
Abstract
A reverse transcription loop-mediated isothermal amplification (RT-LAMP) assay was developed for the rapid detection of equine coronavirus (ECoV). This assay was conducted at 60 °C for 40 min. Specificity of the RT-LAMP assay was confirmed using several equine intestinal and respiratory pathogens in addition to ECoV. The novel assay failed to cross-react with the other pathogens tested, suggesting it is highly specific for ECoV. Using artificially synthesized ECoV RNA, the 50% detection limit of the RT-LAMP assay was 101.8 copies/reaction. This is a 50-fold greater sensitivity than conventional reverse transcription polymerase chain reaction (RT-PCR) assays, but a 4-fold lower sensitivity than quantitative RT-PCR (qRT-PCR) assays. Eighty-two fecal samples collected during ECoV outbreaks were analyzed. ECoV was detected in 59 samples using the RT-LAMP assay, and in 30 and 65 samples using RT-PCR or qRT-PCR assays, respectively. Although the RT-LAMP assay is less sensitive than qRT-PCR techniques, it can be performed without the need for expensive equipment. Thus, the RT-LAMP assay might be suitable for large-scale surveillance and diagnosis of ECoV infection in laboratories with limited resources.
Keywords
Related Topics
Life Sciences
Immunology and Microbiology
Virology
Authors
Manabu Nemoto, Yoshinori Morita, Hidekazu Niwa, Hiroshi Bannai, Koji Tsujimura, Takashi Yamanaka, Takashi Kondo,