Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
6133833 | Journal of Virological Methods | 2014 | 6 Pages |
Abstract
Three types of immunochromatographic assays (ICAs) were designed to detect anti-feline coronavirus (FCoV) antibodies. Recombinant FCoV nucleocapsid protein (rNP) was used as a conjugate or test line in all 3 ICA kits (CJIgG/TNP, CJNP/TNP, and CJNP/TPA). All three ICA kits were capable of detecting anti-FCoV antibodies; however, non-specific positive reactions of anti-FCoV antibody-negative plasma samples with the test line were observed in 2 ICA kits (CJIgG/TNP and CJNP/TNP), in which rNP was used as the test line. On the other hand, the specific detection of anti-FCoV antibodies was possible in all plasma, serum, whole blood, and ascitic fluid samples using the ICA kit with protein A blotted as the test line (CJNP/TPA). In addition, the specificity and sensitivity of ICA (CJNP/TPA) were equivalent to those of the reference ELISA. The development of simple antibody test methods using the principle of ICA (CJNP/TPA) for other coronavirus and feline viral infections is expected in the future.
Related Topics
Life Sciences
Immunology and Microbiology
Virology
Authors
Tomomi Takano, Yuka Ishihara, Masafumi Matsuoka, Shoko Yokota, Yukie Matsuoka-Kobayashi, Tomoyoshi Doki, Tsutomu Hohdatsu,