Abstract

Reverse transcription loop-mediated isothermal amplification (RT-LAMP) was applied to the detection of equine influenza virus (EIV). Because equine influenza is caused currently by EIV of the H3H8 subtype, the RT-LAMP primer set was designed to target the hemagglutinin gene of this subtype. The detection limit of the RT-LAMP assay was a virus dilution of 10−5; which was 103 times more sensitive than the Espline Influenza A&B-N test and 10 times more sensitive than a reverse transcription polymerase chain reaction (RT-PCR) assay. The specificity of the RT-LAMP assay was examined by using several equine pathogens and nasal swabs collected from horses with fever in 2010 after EIV was eradicated in Japan. No cross-reactions were observed. Using 100 nasal swabs collected from horses with fever during an EIV outbreak in 2007, the RT-LAMP assay detected EIV in 52 samples, whereas the Espline test and the RT-PCR assay detected EIV in only 17 and 41 samples, respectively. These results indicate that the RT-LAMP assay is specific for EIV and more sensitive than the Espline test and the RT-PCR assay. Because it provides high sensitivity and ease of manipulation without the need for a thermal cycler or gel electrophoresis, the RT-LAMP assay should be applicable for laboratory diagnosis of EIV.