Article ID Journal Published Year Pages File Type
6134997 Journal of Virological Methods 2010 5 Pages PDF
Abstract
Early diagnosis of HIV infection and appropriate care reduces morbidity and mortality. As a result, recent guidelines recommend that HIV screening be routinely included in patient care. Routine screening will likely result in more patients being tested prior to seroconversion; fourth-generation assays can facilitate diagnosis in these patients. This study evaluated the performance of the automated fourth-generation ADVIA Centaur® HIV Ag/Ab Combo assay. Samples from three sites were tested using the HIV Ag/Ab Combo assay and a CE-marked predicate assay. The HIV Ag/Ab Combo assay's relative sensitivity was 98.36% (599/609; 95% confidence interval: 97.00-99.21%), and the relative specificity was 99.74% (7743/7763; 95% confidence interval: 99.60-99.84%). The HIV Ag/Ab Combo assay detected seroconversion at the same bleed or at least one bleed earlier in 34/37 panels compared to the CE-marked predicate assay. Compared to the final result, the HIV Ag/Ab Combo assay's sensitivity was 100% (598/598; 95% confidence interval: 99.39-100.00%), and the specificity was 99.74% (7753/7773; 95% confidence interval: 99.60-99.84%). Sensitivity was 100% for the HIV genotypes tested. The HIV Ag/Ab Combo assay is a sensitive and specific assay that can assist clinicians in the early diagnosis of HIV infection.
Related Topics
Life Sciences Immunology and Microbiology Virology
Authors
, , , , , ,