Abstract

AimTo evaluate the problems that may arise in breast magnetic resonance imaging (MRI) related to the presence of artefacts and pitfalls, in order to improve its accuracy, sensitivity, and specificity.Materials and methodsSix hundred and thirty breast MRI examinations performed using a 1.5 T magnet were analysed retrospectively. Each type of artefact that may have affected the correct interpretation of the acquired images was considered and analysed. In particular, the presence of technical artefacts, which are related to patient-dependent factors or to the examination itself, and non-technical artefacts, which are the result of inadequate and incorrect image interpretation occurring in absence of technical issues, were examined. In every case of suspicious findings, doubtful lesions were subjected to histological characterization for appropriate therapeutic planning. In the remainder of cases, patients underwent follow-up for at least 18 months.ResultsArtefacts were found in 33% of all examinations, among those 48.6% were caused by movement, 33.6% were due to non-homogeneous or failed fat saturation, 8.7% to incorrect positioning of the patient, 7.2% to metallic artefacts, 1.4% to aliasing, and 0.5% were “zebra artefacts”. When the artefact was identified in a sequence, the sequence was performed a second time after corrective measures. No artefacts affected diagnostic interpretation of the obtained images.ConclusionThe present study provides a specific and precise review of the most frequent artefacts with a discussion of possible and practical solutions. A highly qualified team is required to perform accurate diagnostic tests and to limit or remove the possibility of misinterpretation.

Related Topics
Health Sciences Medicine and Dentistry Oncology
Authors
, , , , , , , , ,