Article ID Journal Published Year Pages File Type
6209662 The Journal of Arthroplasty 2014 4 Pages PDF
Abstract

The stability of prosthetic fixation is to a large extent dependent on component design. The purpose of this study is to analyze the short-term radiological results obtained with the Wagner Standard Cup in primary hip arthroplasty. An assessment was made of one hundred primary hip arthroplasties. The radiological evaluation revealed bone ingrowth in 37 of cases, fibrous integration in 49 and loosening in 14. In summary, osseointegration of the Wagner Standard Cup was unsuccessful in a high percentage of cases. This finding, which was unrelated to the type of stem or bearing surface used, bore a statistically significant relation (P < 0.05) with an observed poor bone coverage of the acetabular component. Although insufficient bone coverage could arguably be attributable to a poor surgical technique, we would tend to agree with other authors that it is rather likely to result from a flawed implant design that impedes osseointegration.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, , , , , , ,