Article ID Journal Published Year Pages File Type
6210001 The Journal of Arthroplasty 2013 6 Pages PDF
Abstract

Accurate implant positioning and restoration of lower limb alignment are major requirements for successful long-term results in unicompartmental knee arthroplasty (UKA). Alignment accuracy was compared between navigated-UKA (nUKA) and conventional-UKA (cUKA) groups using a retrospective matched case-control study (n = 129, 58 nUKA, 71 cUKA). Mechanical axis (MA), hip-knee-ankle angle (HKA°), coronal implant alignment, and tibial implant posterior slope were measured. No statistically significant difference was observed when comparing MA, HKA° or coronal implant alignment (p > 0.05). Statistical significance was seen with tibial component posterior slope (p = 0.04, nUKA 4.2°, cUKA 2.9°); and between intra-operative navigationally determined HKA° and post-operative whole-leg standing (WLS) film HKA°. Navigation does not significantly improve UKA alignment compared to conventional methods. Further studies are needed to justify the use of this technology in UKA.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, , , , ,