Article ID Journal Published Year Pages File Type
6211537 The Knee 2009 4 Pages PDF
Abstract

The St. Leger total knee replacement (Zynergy Orthopaedics Ltd, Rotherham, UK) was developed as a cheaper alternative to similar implants of its time. Between October 1993 and June 1999, 144 St. Leger total knee replacements were implanted into 114 patients. Seventy-three patients (99 knees) were recalled for assessment (mean follow-up of 10.2 years). Eighteen patients had had their prostheses revised, 11 had died and 12 were lost to follow-up. Functional Score showed 90% poor results and the Objective Knee Score showed 31% poor results. Radiological assessment identified 12 arthroplasties that had failed and 58 that required close follow. Kaplan-Meier cumulative survivorship was 87% at 10 years. The St. Leger knee replacement did not perform as well as others of the same generation and was not worth the initial financial savings.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, , , ,