Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
6350804 | Environmental Research | 2017 | 7 Pages |
Abstract
The aims of this study were to characterize electromagnetic fields of radiofrequency (RF-EMF) levels generated in a Neonatal Medium Care Unit and to analyze RF-EMF levels inside unit's incubators. Spot and long-term measurements were made with a dosimeter. The spot measurement mean was 1.51±0.48 V/m. Higher values were found in the proximity to the window and to the incubator evaluated. Mean field strength for the entire period of 17 h was 0.81 (±0.07) V/m and the maximum value was 1.58 V/m for long-term RF-EMF measurements in the incubator. Values found during the night period were higher than those found during the day period. It is important to consider RF-EMF exposure levels in neonatal care units, due to some evidence of adverse health effects found in children and adults. Characterization of RF-EMF exposure may be important to further investigate the mechanisms and underlying effects of electromagnetic fields (EMF) on infant health. A prudent avoidance strategy should be adopted because newborns are at a vulnerable stage of development and the actual impact of EMF on premature infants is unknown.
Related Topics
Life Sciences
Environmental Science
Health, Toxicology and Mutagenesis
Authors
I. Calvente, A. Vázquez-Pérez, M.F. Fernández, M.I. Núñez, A. Múñoz-Hoyos,