Article ID Journal Published Year Pages File Type
6552327 Forensic Science International 2014 32 Pages PDF
Abstract
Five hundred microliters of whole blood were extracted by routine solid-phase extraction (SPE, HCX). Analysis was performed on an ABSciex 3200 QTrap instrument with ESI+ in scheduled MRM mode. The method was fully validated according to international guidelines including selectivity, recovery, matrix effects, accuracy and precision, stabilities, and limit of quantification. The selected SPE provided recoveries >60% for all analytes except 6-monoacetylmorphine (MAM) with coefficients of variation (CV) below 15% or 20% for quality controls (QC) LOW and HIGH, respectively. Ion suppression >30% was found for benzoylecgonine, hydrocodone, hydromorphone, MDA, oxycodone, and oxymorphone at QC LOW, however CVs were always below 10% (n = 6 different whole blood samples). Accuracy and precision criteria were fulfilled for all analytes except for MAM. Systematic investigation of accuracy determined for QC MED in a multi-analyte mixture compared to samples containing only single analytes revealed no relevant differences for any analyte, indicating that a multi-analyte calibration is suitable for the presented method. Comparison of approximately 60 samples to a former GC-MS method showed good correlation. The newly validated method was successfully applied to more than 1600 routine samples and 3 proficiency tests.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry