Abstract
Methodology is presented for the profiling of fentanyl in seized drugs using ultra high performance liquid chromatography in combination with tandem mass spectrometry (UHPLC-MS/MS). Target analysis was performed for 40 fentanyl processing impurities, several of which are markers for a specific synthetic route (Siegfried or Janssen). For the separation of these solutes, an Acquity BEH C18 1.7 μm particle column (150 mm × 2.1 mm) with a binary 1% formic acid (pH 2.0)/acetonitrile gradient was used. For MS/MS detection, an atmospheric pressure positive electrospray source was employed with selected reaction monitoring (SRM). The coupling of the high separation power of UHPLC with the highly selective and sensitive detection of MS/MS is amenable to the determination of synthetic route and linking of drug seizures. The technology is also applicable to exhibits containing trace levels of fentanyl in the presence of significantly excess amounts of heroin and/or adulterants.