Abstract
The Investigator® Quantiplex HYres kit was evaluated as a potential replacement for dual DNA quantification of casework samples. This kit was determined to be highly sensitive with a limit of quantification and limit of detection of 0.0049 ng/μL and 0.0003 ng/μL, respectively, for both human and male DNA, using full or half reaction volumes. It was also accurate in assessing the amount of male DNA present in 96 mock and actual casework male:female mixtures (various ratios) processed in this exercise. The close correlation between the male/human DNA ratios expressed in percentages derived from the Investigator® Quantiplex HYres quantification results and the male DNA proportion calculated in mixed AmpFlSTR® Profiler® Plus or AmpFlSTR® Identifiler® Plus profiles, using the Amelogenin Y peak and STR loci, allowed guidelines to be developed to facilitate decisions regarding when to submit samples to Y-STR rather than autosomal STR profiling. The internal control (IC) target was shown to be more sensitive to inhibitors compared to the human and male DNA targets included in the Investigator® Quantiplex HYres kit serving as a good quality assessor of DNA extracts. The new kit met our criteria of enhanced sensitivity, accuracy, consistency, reliability and robustness for casework DNA quantification.
Related Topics
Life Sciences
Biochemistry, Genetics and Molecular Biology
Genetics
Authors
Chantal J. Frégeau, Nancy Laurin,