Article ID Journal Published Year Pages File Type
6662240 Journal of Electroanalytical Chemistry 2017 27 Pages PDF
Abstract
Furthermore, silver electrode can be used successfully to improve the detection of 4-NA at very low concentration levels with high sensitivity. The linear calibration range was from 8.0 × 10− 9 to 1.0 × 10− 3 mol L− 1. The limit of detection (LOD) was found to be 4.74 × 10− 9 mol L− 1, while relative standard deviation (RSD) at 1.0 × 10− 4 mol L− 1 4-NA concentration was 1.94% for eight repetitions. Then, the proposed method was applied to detect 4-NA in wastewater and drinking water with satisfactory results.
Related Topics
Physical Sciences and Engineering Chemical Engineering Chemical Engineering (General)
Authors
, , , , , ,