Article ID Journal Published Year Pages File Type
7124168 Measurement 2016 8 Pages PDF
Abstract
The aim of this study is to assess the sensitivity of MRI thermometry using three T1-weighted sequences (i.e., Inversion Recovery Turbo-FLASH, IRTF, Saturation Recovery Turbo-FLASH, SRTF, and FLASH) using an 1.5-T MR scanner on healthy swine pancreases undergoing LA. The reference temperature was measured by MRI-compatible fiber optic sensors (fiber Bragg grating sensors). The sensitivity of the proposed techniques was estimated and compared. The thermal sensitivity of the three sequences was −1.47 ± 0.08 °C−1, −0.95 ± 0.05 °C−1, and −0.56 ± 0.04 °C−1 for IRTF, SRTF and FLASH, respectively. Results show that the proposed technique may be adequate for temperature monitoring during LA.
Related Topics
Physical Sciences and Engineering Engineering Control and Systems Engineering
Authors
, , , , , , ,