Article ID Journal Published Year Pages File Type
7207130 Journal of the Mechanical Behavior of Biomedical Materials 2018 10 Pages PDF
Abstract
The pull-out forces of the Ti-variant (complete-526 N, reduced-468 N) and the Ti-S variant (complete-548 N, reduced-526 N) as well as the lever-out moments of the Ti-variant (complete-10 Nm, reduced-9.8 Nm) and the Ti-S variant (complete-9 Nm, reduced-7.9 N) show no significant differences in the results between complete and reduced cups. The results show that the use of reduced cups in a press-fit design is possible within the scope of development work.
Related Topics
Physical Sciences and Engineering Engineering Biomedical Engineering
Authors
, , , ,