Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
7207130 | Journal of the Mechanical Behavior of Biomedical Materials | 2018 | 10 Pages |
Abstract
The pull-out forces of the Ti-variant (complete-526 N, reduced-468 N) and the Ti-S variant (complete-548 N, reduced-526 N) as well as the lever-out moments of the Ti-variant (complete-10 Nm, reduced-9.8 Nm) and the Ti-S variant (complete-9 Nm, reduced-7.9 N) show no significant differences in the results between complete and reduced cups. The results show that the use of reduced cups in a press-fit design is possible within the scope of development work.
Related Topics
Physical Sciences and Engineering
Engineering
Biomedical Engineering
Authors
Volker WeiÃmann, Christian Boss, Rainer Bader, Harald Hansmann,