Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
7207803 | Journal of the Mechanical Behavior of Biomedical Materials | 2016 | 4 Pages |
Abstract
In this technical note, a simple theoretical model was proposed to express the tensile deformation and fracture of balloon catheter tested by the ISO standard using piece-wise linear force-displacement relations. The model was then validated by comparing with the tensile force-displacement behaviors of two types of typical balloon catheters clinically used worldwide. It was shown that the proposed model can effectively be used to express the tensile deformation behavior and easily be handled by physicians who are not familiar with mechanics of materials.
Related Topics
Physical Sciences and Engineering
Engineering
Biomedical Engineering
Authors
Mitsugu Todo, Keiji Yoshiya, Takuya Matsumoto,