Article ID Journal Published Year Pages File Type
7209258 Journal of the Mechanical Behavior of Biomedical Materials 2013 10 Pages PDF
Abstract
The elastic properties of the vocal folds (VFs) vary as a function of depth relative to the epithelial surface. The poroelastic anisotropic properties of porcine VFs, at various depths, were measured using atomic force microscopy (AFM)-based indentation. The minimum tip diameter to effectively capture the local properties was found to be 25 µm, based on nonlinear laser scanning microscopy data and image analysis. The effects of AFM tip dimensions and AFM cantilever stiffness were systematically investigated. The indentation tests were performed along the sagittal and coronal planes for an evaluation of the VF anisotropy. Hertzian contact theory was used along with the governing equations of linear poroelasticity to calculate the diffusivity coefficient of the tissue from AFM indentation creep testing. The permeability coefficient of the porcine VF was found to be 1.80 ± 0.32 ×10−15 m4/N s.
Related Topics
Physical Sciences and Engineering Engineering Biomedical Engineering
Authors
, , , , ,