| Article ID | Journal | Published Year | Pages | File Type |
|---|---|---|---|---|
| 7211592 | Ciência & Tecnologia dos Materiais | 2017 | 4 Pages |
Abstract
Three brands of rotary nickel-titanium endodontic instruments with some comparable geometric features but with different cross-section (similar tip size and same taper of .04) were selected for this study: MTwo .04(35) (VDW GmbH, Germany), K3 .04(30) (SybronEndo, Mexico) and K3XF .04(30) (SybronEndo, Mexico). K3XF is made from novel R-phase heat-treated metal alloy while all other files were made from traditional Ni-Ti alloy. The instruments were analysed under the following conditions: i) as-received (AR), ii) heat treated at 350 °C, iii) heat treated at 400 °C. The transformation temperatures were determined by differential scanning calorimeter (DSC). Compared to conventionally K3 and MTwo files, R-phase heat treatment K3XF file showed higher transformation temperatures. These results also showed that heat treatments increase the transformation temperatures.
Related Topics
Physical Sciences and Engineering
Engineering
Engineering (General)
Authors
F.M. Braz Fernandes, A.R. Alves, A. Machado, J.P. Oliveira,